SECTION 16 b. TROUBLE SHOOTING CONT.
available. If the 2nd module functions properly consult the factory
about the faulty module if both modules fail consult the factory
regarding the HHC base unit.
WILL NOT DISPLAY TEMPERATURE MODULE DATA
For both Thermocouple and RTD modules, ensure that the
module is connected to either a thermocouple or RTD probe and
that the sections of the manual relating to their set up has been
observed .
Follow the procedure for lack of pressure module readings under
“HHC WILL NOT DISPLAY PRESSURE MODULE DATA” above
SYSTEM ERROR MESSAGES
Internal Operating System Error Messages
If any of the following messages are displayed Reset the power
off and back on.
If cycling power does not resolve the issue, note the sequence
of operations performed before the error message appeared and
contact customer service.
Error message list:
Operation timed out triggering self-protection reset
Failed to create internal message queue
Nonexistent notice or error message received
A system error occurred, and the error message was
corrupted
Failed to initialize display subsystem
Failed to initialize ADC subsystem
Failed to initialize module subsystem
Failed to initialize keypad subsystem
Failed to initialize periodic checks subsystem
Failed to initialize director subsystem
Failed to initialize MENU subsystem
Failed to initialize communications subsystem
Failed to initialize datalog subsystem
Stack overflow in error handling subsystem
Stack overflow in display subsystem
Stack overflow in ADC subsystem
Stack overflow in module subsystem
Stack overflow in keypad subsystem
Stack overflow in periodic checks subsystem
Stack overflow in director subsystem
Stack overflow in MENU subsystem
Stack overflow in communications subsystem
Stack overflow in datalog subsystem
Stack overflow in unknown subsystem
Internal Hardware System Errors
Internal communication failure (I2C bus)
Communication failure with module EEPROM
Corrective action: Confirm Module(s) are inserted properly.
Reset power off/on;
If problem persists contact customer service
ADC reference not within valid range
ADC conversion timed out
Corrective action: Reset power off/on, contact customer service
System EEPROM appears to be corrupted
Corrective action: Reset power off/on, contact customer service
Hardware Warnings
CH1 F.S. INVALID/ CH2 F.S. INVALID: