Quick Start Function Summary
Instructions for ASHCROFT
®
GC51
This function allows the user to establish a linear relation-
ship from the standard “psi” unit to any user defined unit.
Note: See menu schematic at end, must be in “Linear
Display Mode” option within “Setting Mode”, this is
noted on the screen by
Use
$#
keys to move between “Linear Display Mode”
and “Pressure Display Mode.
Note: Values shows are from example in I&M manual.
© 2008 Ashcroft Inc., 250 East Main St., Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0499, www.ashcroft.com
All sales subject to standard terms and conditions of sale. 06/09 I&M011-10063-GC51
Note: Holding the button for more than
3
seconds returns to display mode.
Complete Setting Mode Menu
n L;n
d 0.00
d 5.00
d 2
p 60.0
a 100.0
a 0.0
p 10.0
-
Setting LCD Setting Setting
Item Display Description Range
Display Selection of non:pressure
mode pressure display display mode
mode : Lin Lin:Linear display
mode
Minimum Min. pressure Pressure range
pressure corresponding to 0 to 75% Span
OFFSET 9
: 10 (psi)
Maximum Max. pressure Pressure range
pressure corresponding to 25 to 100% Span
FULL SCALE 10
: 60 (psi)
Decimal Display after 0,1,2,3 digit
point decimal point
position Number of digits
: 2 (digit)
OFFSET OFFSET –1999 to 1999
corresponding to
min. pressure 6
: 0.00 (ton)
FULL FULL SCALE –1999 to 1999
SCALE corresponding to
mAX. pressure 7
: 5.00 (ton)
Output Analog output Max. display
zero point zero point (4mA) span: –10
: 0.0 ( %Span) to 110% Span
Output Analog output Max. display
span point span point (20mA) span: –10
: 100.0 (%Span) to 110% Span
D. Loop Check: Use to send a 4-20mA signal meant to
a 90.0
a 10.0
n non
-
Note: For setting of zero point and span point in the analog
o
utput, input the percent value over the pressure range.
Setting LCD Setting Setting
Item Display Description Range
Display Selection of non:pressure
m
ode pressure display display mode
mode : non Lin: linear
display mode
Output Analog Pressure
zero point output zero point range:–10
(
4mA) : 10.0 (%Span) to 110% Span
.
Output Analog Pressure
span point output span point range:–10
(20mA) : 90.0 (%Span) to 110% Span
n L;n
-
simulate applied pressure, can be accessed either
through Pressure Display Mode or Linear Display Mode.
See “Complete Setting Mode Menu” at end. Loop Check
is noted on the screen with a prefix
(
. The display
is indicating in actual units and starts at the zero
(4mA) point.
If
$
button continues to be pressed, the linear display
will auto-increment by linkage between the linear display
and the analog output. By continuing to press
#
button,
auto decrement will occur. Release the button at the
desired indication.
Notes shown below are from I&M manual.
Notes: Actual values shown are based upon the examples
shown in the I&M Manual.
Changes made within the Setting Mode are saved
by returning to Measurement Mode before powering
the unit “off.