13.2 Setting Items for Linear Display Mode (Re-scaling in arbitrary user
defined units) Set the filter before setting the linear display mode
(Refer to the preceding Section 11.1). The setting of the following table
is the Setting example 2: Linear display mode of Section 11.3.
(Arbitrary unit: m). This applies when re-scalingin arbitrary user
defined units. See page 24 for full menu.
No Setting Item LCD Display Setting Description Setting Range Default
n00l0n
Selection of linear
display mode: Lin
non: Differential
pressure display
mode; Lin: Linear
display mode
Display mode
-
-
0.0
p0 -20
Min. differential
pressure corre-
sponding to OFF-
SET
:20.0(inH
2
O)
Differential pres-
sure range:
0 to 75%F.S.
Min. differen-
tial pressure
(1)
100.0
Max. differential
pressure corre-
sponding to FULL
SCALE
:120inH
2
O)
Differential
pressure range:
25 to 100%F.S.
Max. differen-
tial pressure
(1)
p0-120
0
Display after deci-
mal point Number
of digits:1(digit)
0,1,2,3 digit
Decimal point
position
d0-001
d0-00.0
d0-50.0
0
OFFSET correspon-
ding min. differential
pressure
: 0.0 (m)
–1999 to 1999
OFFSET
1000
FULL SCALE corre-
sponding to max.
differential pres-
sure
:50.0 (m)
–1999 to 1999
FULL SCALE
a0-50.0
0.0
Analog output zero
point : (4mA): 0.0
(%F.S.)
Max. display span:
–10 to 110%F.S.
Output zero
point
(2)
a0100.0
100.0
Analog output span
point : (20mA):
100.0 (%F.S.)
Max. display span:
–10 to 110%F.S.
Output span
point
(2)
[0-50.0
0
(4.0mA)
Arbitrary change of
linear display and
analog output:
50.0 (m), 20mA
Display: Linear dis-
play span; Analog
output: 4 to 20mA
Loop check
(3)